SC12 Magnetic Field Cancelling System |
|
Protecting your Investment in Electron Beam Technology |
|
SC12 Product Specification
|
|
COORDINATE SYSTEM
|
X, Y, Z rectangular Cartesian |
| |
|
UNITS
|
Gauss, Tesla (switchable) |
| |
|
FIELD CANCELLING
|
|
| |
1. |
Standard SC12 system
|
|
| |
|
Components cancelled
|
X, Y, Z field components |
| |
|
Dynamic range
|
50 milliGauss pk-pk per axis |
| |
|
Field reduction ratio
|
50X (typical) at 50/60Hz |
| |
|
Bandwidth
|
0.5Hz - 5000Hz |
| |
|
Detector noise limit
|
0.002 milliGauss pk-pk |
| |
| |
2. |
Wideband DC sensor system
|
|
| |
|
Components cancelled
|
X, Y, Z field components |
| |
|
Ambient DC field
|
2 Gauss, Max |
| |
|
Dynamic range
|
40 milliGauss pk-pk per axis |
| |
|
Field reduction ratio
|
50X (typical) at 50/60Hz |
| |
50X (typical) at DC (incremental) |
| |
|
Bandwidth
|
DC - 2.5kHz |
| |
|
Detector noise limit
|
DC: 30µG Pk-Pk typ. (0.0001 - 0.01 Hz) |
| |
AC: 1µG/√Hz RMS typ. at 50/60Hz |
| |
|
DC drift
|
<100 µG/24 hours |
| |
|
FIELD MEASUREMENT
|
|
| |
Types
|
1. Real time field |
| |
2. AC - true RMS amplitude |
| |
3. Incremental DC (with DC sensor) |
| |
| |
Display
|
|
| |
|
RMS & DC
|
3.5 digit LCD panel meter |
| |
|
Sensor dynamic range
|
40 milliGauss pk-pk |
| |
|
Meter ranges
|
| X1 range, | 0-1.999mG RMS |
| +/-1.999mG DC |
| X10 range, | 0-19.99mG RMS |
| +/-19.99mG DC |
| X100 range, | 0-199.9mG RMS |
| +/-199.9mG DC |
| (plus equivalent Tesla ranges) |
|
| |
| |
Accuracy
|
|
| |
|
RMS
|
+/-1.0% of reading +/0.003mG |
| |
|
Incremental DC
|
±5% (after >1hour warm up) |
| |
| |
Real time field monitor
|
|
| |
|
Signals
|
X, Y, Z real time fields |
| |
|
Scaling
|
500mV/mG |
| |
|
Range
|
+/-10 Volts |
| |
|
Source resistance
|
10K |
| |
|
Connectors
|
3 X BNC |
| |
|
Bandwidth
|
| with AC sensor: | 5Hz - 20kHz |
| with DC sensor: | DC - 5kHz |
|
| |
| |
Trip range adjustment
|
|
| |
|
AC-RMS
|
0 - 2.0 mG |
| |
|
DC
|
0 - 5.0 mG |
| |
|
POWER
|
120V 60Hz or 240V 50Hz, 50VA |
|